Delivery modes
|
Continuous, Bolus only, Continuous + Bolus, Programmed Automatic Bolus (PAB)
|
Accuracy
|
+/- 5 %
|
Flow rate
|
0.1 to 100 ml/h
|
Programming units
Concentration
|
ml/mg/µg
0.1 to 99.9 mg/ml/0.1 to 999.9 µg/ml
|
Volume to infuse
|
1 to 9999 ml
|
Bolus dose
|
0 or from 0.1 to 60ml with 0.1ml increment
|
Bolus flow rate
|
100 ml/h, adjustable from 50 to 100 ml/h with 10 ml/h increment
|
Lockout interval
|
0 to 999 minutes
|
Special functions
|
Protocol libraries (60 protocol per library)
Lockable pump keyboard
Autopriming through the pump
Dose limit / x hours and Bolus limit / x hours
Loading dose
Clinician bolus override
Mandatory priming (enabled, warning only, disabled)
|
Therapy monitoring
|
Recording of patient pain scores with visual analogue scale (VAS)
VAS available in the pump history, printable and computer downloadable
"Ready to Connect" to the MicrelCare System for therapy monitoring through the web.
|
KVO
|
0.1 to 2.0 ml/h
|
Automatic restart of infusion after occlusion release
|
Yes
|
Safety features
|
Air in line eliminating filter
Lockable protective cover for drug container
Code protected stop of infusion
Free flow protection
Programming within protocol limits
Dedicated therapy front face for regional analgesia (PCEA/PCRA yellow color code)
|
Occlusion alarm levels
|
Low (0.8 +/- 0.4 = 0.4 - 1.2 bar)
Middle (1.2 +/-0.4 = 0.8 - 1.6 bar)
High (1.6 +/- 0.4 = 1.2 - 2.0 bar)
|
Alarms & warnings
|
Occlusion near end and end of infusion
Keyboard continuously pressed
Low and depleted battery
Dose limit, bolus limit
Open mechanism, preventive maintenance warnig
|
Access code levels
|
3
|
Programming
|
Protocol libraries programmed with "Rythmic™ Evolution Therapy Manager" computer software
Direct programming on the pump possible with code
|
History, infusion graph / printout and PC download
|
Yes, including 6500 time stamped events event log
Infusion graph, bolus graph/list, pain level/graph/list
"Plug and play" print facility
|
Battery life
|
9 days at 2 ml/h
|
Power supply
|
100 to 240 V, 50-60 Hz (with optional adaptor KS5.01.615.4)
|
Weight with battery
|
300g for Mini Rythmic™ Evolution
335g for Rythmic™ Evolution
740g for Rythmic™ Organizer 500 series
|
Size
|
130mm x 75mm x 46mm for Mini Rythmic™ Evolution (including 100 ml bag)
130mm x 133mm x 46mm for Rythmic™ Evolution (including 200 ml bag)
157mm x 299mm x 82mm for Rythmic™ Organiser 500 series (including 500 ml standard bag)
|