mini Rythmic™ PN |
 |
|
|
|
|
An ambulatory pump for Parenteral Nutrition
and continuous therapies, the mini Rythmic PN has been
designed:
- To use external bag
to infuse unlimited volume taking advantage of a compact
pump design.
- To incorporate air
filter eliminator upstream the air detector to avoid
repeated air alarms and related priming constraints.
- To increase operational autonomy and reduce carried
weight.
With infusion rates from 0.1 - 400.0
ml/hr (0.1 to 999.9 mg/hr), the mini Rythmic PN can be used for
any continuous high demand treatment such as Parenteral
Nutrition, chemotherapy, and antibiotic therapies.
|
|
|
|
mini Rythmic™ PN
is "Alarm free" |
 |
|
|
|
|
|
|
|
|
|
|
|
|
| |
The pump has been designed to decrease most
of unnecessary potential alarms whilst preserving full safety:
-
No air alarm thanks to the air in line eliminator, positioned upstream the AIL detector, so that air alarm means defective filter
-
Automatic
restart after occlusion release.
-
Long battery autonomy ( 17h @300ml/h
with lightweight battery pack)
to decrease the frequency of battery alarms and related
changes.
|

|
|
|
|
mini Rythmic™ PN offers "All in one"
innovative and safety infusion sets |
 |
|
| |
-
All sets include all pre-assembled components thus avoiding
additional unsafe connections.
-
In built anti free flow / back check valve, air eliminator,
and extra long extension
line on all PN sets.
Visit our disposables
page, view PN infusion sets |
|
|
|
mini Rythmic™ PN
offers real mobility
to patients |
 |
|
| |
-
Small, ergonomic and compact for true and comfortable
ambulatory therapies
-
Extensive range of rucksacks (photo of 4L adult one)
-
Lightweight: 290g with dry cell battery
-
Outstanding battery autonomy and small size (2x5.5x10cm) and weight (100g) battery pack
The pump is also supported by a wide range of
accessories. |

|
|
|
|
mini Rythmic™ PN
Downloads |
 |
|
| |
Rythmic brochure 05 2007.pdf (1.00 MB)
Rythmic brochure 05 2007.pps (2.75 MB)
Rucksack
DFU 4L adult
Rucksack
DFU 2L pediatric/2 PUMPS
Declaration of Conformity for
Mini Rythmic PN
Approvals (SGS UK Ltd)
|
|
|
|
mini Rythmic™ PN
Specifications |
 |
|
|
|
Delivery Modes |
Continuous, Volume /time, Volume/rate, auto ramp |
|
Accuracy |
+/- 5% |
|
Flow rate |
0.1 to 400ml/h |
|
Programming units Concentration |
ml/mg/ėg , 0.1 to 99.9 mg/ml, 0.1 to 999.9 ėg/ml |
|
Volume to infuse |
1 to 9999ml |
|
Programmable infusion time |
10min to 99 hours |
|
Auto ramp, up & down time |
1min to 9h59 |
|
Early taper down |
yes |
|
Delayed start |
0 to 24h |
|
flat rate titration |
yes |
|
Air bubble size (to detect air eliminating filter
failure) |
0.1ml; 0.5ml; 2ml; off |
|
Automatic restart after occlusion release |
yes |
|
KVO |
0.1 to 20.0ml/h |
|
AC input |
11VDC 500mA |
|
Lock level |
3 |
|
Historic |
1500 time stamped events |
|
Historic graph |
volume/ time |
|
Connectivity |
portable printer, PC |
|
Alarms |
Occlusion, near end and end of infusion, low and
depleted battery, open mechanism, preventive maintenance
warning, dose limit, bolus limit, air eliminating filter
defect (normal use has no air alarms) |
|
Safety features |
Air in line eliminator, free flow protection according
to IEC 601-2-24 |
|
Battery autonomy with 9V alkaline |
with 9V alkaline: 5 h @300ml/h or 3,5L @400ml/h; |
|
Battery autonomy with battery pack |
17h @300ml/h |
|
Weight with battery |
290g |
|
size incorporating air eliminating filter |
80mmx130mmx46mm |
|
|
|
mini Rythmic™ PN Extensive Accessories list |
 |
|
|
|
|
|
|
|
|
|
|
|
|
- Recheargeable Battery pack very small
size
- 2L, 4L,
single, dual or triple pump rucksacks
-
Pole clamp
-
Power Adaptor / Battery Pack
charger also as UK version
-
12V car plug adaptor
- Portable events printer
|
|
|